Skip to main content

Network interfaces

Each CID has a Corporate network interface and an Instrument network interface. Their separate roles keep instrument traffic isolated while allowing the CID to reach CID Hub, OpenLab services, and CDS clients. Both interfaces support automatic addressing through Dynamic Host Configuration Protocol (DHCP) or a static configuration.

Interface roles​

The two interfaces serve different networks and must not share a default route.

InterfaceRoleTypical configuration
Corporate NICConnects the CID to the corporate network, CID Hub, the OpenLab Server, and CDS clients.DHCP is required for initial activation. Automatic configuration is recommended; static configuration is available after activation.
Instrument NICConnects the Windows VM to an isolated instrument network.Static address in the instrument subnet; no default gateway and normally no DNS server.
warning

Leave Gateway Address (Not Recommended) blank on the Instrument NIC. A gateway can create a competing default route or allow instrument traffic to cross network boundaries.

Configuration fields​

CID Hub displays these values for each interface.

FieldMeaning
Configuration methodAutomatic addressing or manually supplied values.
Connection stateWhether the operating system reports the interface as connected.
IP AddressThe IPv4 address assigned to the interface.
Subnet MaskThe IPv4 subnet mask.
MAC AddressThe physical interface address.
Gateway AddressThe route used to reach networks outside the local subnet. Use this only on the Corporate NIC.
DNS AddressOne or more DNS resolvers. These are normally required only on the Corporate NIC.
Domain NameThe DNS search domain reported by the interface.

For requirements, see System requirements. For configuration instructions, see Configure network cards.